GE LOGIQ Totus Basic Service Manual page 558

Table of Contents

Advertisement

D
5936427, R
IRECTION
EV
8-12-62
Replacement of the COM EXPRESS MODULE, I5 CFL
Functional Check
4-2-3 "Power ON/Boot Up"
4-3-8 "B-Mode functional checks"
4-3-20 "Basic measurement functional checks"
10-6-4 "Grounding Continuity"
10-6-6 "Earth Leakage Current Test"
4-2-4 "Power shut down"
8-12-63
Replacement of the GPU SET-T
Functional Check
4-2-3 "Power ON/Boot Up"
4-3-8 "B-Mode functional checks"
4-3-20 "Basic measurement functional checks"
10-6-4 "Grounding Continuity"
10-6-6 "Earth Leakage Current Test"
4-2-4 "Power shut down"
8-12-64
Replacement of the UEPP-T
Functional Check
4-2-3 "Power ON/Boot Up"
4-3-8 "B-Mode functional checks"
4-3-20 "Basic measurement functional checks"
10-6-4 "Grounding Continuity"
10-6-6 "Earth Leakage Current Test"
4-2-4 "Power shut down"
8-12-65
Replacement of the M.2 SSD 1TB
Functional Check
4-2-3 "Power ON/Boot Up"
4-3-8 "B-Mode functional checks"
4-3-20 "Basic measurement functional checks"
10-6-4 "Grounding Continuity"
10-6-6 "Earth Leakage Current Test"
4-2-4 "Power shut down"
8 - 232
Section 8-12 - Functional Checks after FRU replacement (Debrief Guideline)
GE HealthCare CONFIDENTIAL
. 1
Debrief
Script
Equipment passed all required checks and is ready for
use.
Debrief
Script
Equipment passed all required checks and is ready for
use.
Debrief
Script
Equipment passed all required checks and is ready for
use.
Debrief
Script
Equipment passed all required checks and is ready for
use.
LOGIQ T
Basic Service Manual
OTUS
16G, I5 CFL 16