Monitor Adjustments - GE LOGIQ Totus Basic Service Manual

Table of Contents

Advertisement

Direction 5936427, Rev. 1
6-2-3

Monitor adjustments

Room profile
To set the monitor to different room lighting conditions, you may want to manually adjust the room
profile.
To set the room profile:
1.) Navigate to Utility (second page) > Scan Screen. The scan screen displays on the main display
while the Utility Touch Panel is active so you can the affect your adjustments are making to the
monitor.
2.) Under Room Profile, select the room type.
Table 6-2
Room Profile
Semi-Dark
User Defined
Backlight
Under special lighting conditions, you may want to adjust the monitor screen's backlight intensity.
To set the OPIO backlight:
1.) Navigate to Utility (second page) > Scan Screen. The scan screen displays on the main display
while the Utility Touch Panel is active so you can the affect your adjustments are making to the
monitor.
2.) Under the operator panel, turn the encoder to adjust the OPIO backlight setting.
Brightness and contrast
Under special lighting conditions, you may want to manually adjust the brightness and contrast.
To manually set the brightness and contrast:
1.) Navigate to Utility (second page) > Scan Screen. The scan screen displays on the main display
while the Utility Touch Panel is active so you can the affect your adjustments are making to the
monitor.
2.) Under Room Profile, select User Defined.
3.) Under the operator panel, turn the encoder to adjust the brightness/contrast setting.
6 - 4
GE HealthCare CONFIDENTIAL
Room Profile Table
Brightness/Contrast HDU
Dark
Light
User-adjustable Brightness/Contrast
Section 6-2 - Monitor Adjustments
52/50
60/70
62/93
LOGIQ Totus Basic Service Manual
Brightness/Contrast LCD
50/85
70/85
90/85
User-adjustable Brightness/Contrast

Advertisement

Table of Contents
loading

Table of Contents