8-7-12 Replacement Of The T2 Opio Cable Hook - GE LOGIQ Totus Basic Service Manual

Table of Contents

Advertisement

D
5936427, R
IRECTION
EV

8-7-12 Replacement of the T2 OPIO Cable Hook

This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
Table 8-48
Manpower
One person
8-7-12-1
Removal Procedure
Unscrew 2 screws on the left or right side of the T2 OPIO Cable
1.
Hook to remove from the console.
8-7-12-2
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform "Functional Checks after FRU replacement (Debrief Guideline)" as described in
12 on page 8-223
8 - 102
GE HealthCare CONFIDENTIAL
. 1
Manpower, Time, Tools, and PPE
Time
Approximately 5 minutes
Step
Section 8-7 - Replacing OPIO
LOGIQ T
Tools
Phillips Screwdriver
Corresponding graphic
Basic Service Manual
OTUS
8-2-6 "Tools needed
PPE
Cut resistant gloves
Knee Pad
Section 8-