Direction 5936427, Rev. 1
10-6-6
Earth Leakage Current Test
10-6-6-1
Definition
This test measures the current that would flow to a grounded person who touched accessible
protectively-earthed parts of equipment if the ground wire should break.
The meter is connected from accessible protectively-earthed parts of the equipment to ground.
Measurements should be made with power line polarity normal and reversed, and with the neutral open
and closed Record the highest reading.
Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit
CAUTION
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged.
!! CAUTION:
GE HealthCare CONFIDENTIAL
Chapter 10 Care & Maintenance
LOGIQ Totus Basic Service Manual
10 - 29