GE LOGIQ Totus Basic Service Manual page 467

Table of Contents

Advertisement

D
5936427, R
IRECTION
EV
8-8-13 Replacement of the Castor Dual Lock N9
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
Table 8-75
Manpower
Two people
8-8-13-1
Preparations
1.) Remove the Unity Base Tray as described in
8-130
2.) Remove the Front Cover Assy-T as described in
page 8-131
3.) Remove the T2 Side Cover L Assy as described in
Assy" on page 8-135
4.) Remove the T2 Side Cover R Assy as described in
Assy" on page 8-136
5.) Remove the Rear Cover Mid Assy-T as described in
Assy-T" on page 8-138
6.) Remove the Base Cover Set-T as described in
page 8-140
8-8-13-2
Removal Procedure
Unscrew the nut to remove the Castor Dual Lock
1.
N9 from the console.
Press the caster wheel locks and ensure that they keep the system from moving.
NOTICE
Proceed with extreme caution and avoid system tipping over. The system weighs
WARNING
WARNING
approximately 85kg.
8-8-13-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform "Functional Checks after FRU replacement (Debrief Guideline)" as described in
12 on page 8-223
GE HealthCare CONFIDENTIAL
. 1
Manpower, Time, Tools, and PPE
Time
Approximately 15 minutes
Step
Section 8-8 - Replacing Main Console
LOGIQ T
Tools
24mm Socket or
Adjustable wrench
8-8-2 "Replacement of the Unity Base Tray" on page
8-8-3 "Replacement of the Front Cover Assy-T" on
8-8-7 "Replacement of the T2 Side Cover L
8-8-8 "Replacement of the T2 Side Cover R
8-8-10 "Replacement of the Rear Cover Mid
8-8-12 "Replacement of the Base Cover Set-T" on
Corresponding graphic
Basic Service Manual
OTUS
8-2-6 "Tools needed
PPE
Cut resistant gloves
Knee Pad
Section 8-
8 - 141