8-7-16 Replacement of the OPIO Knob Set-T

This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
Table 8-52
Manpower
One person
8-7-16-1
Removal Procedure
1.
Pull 13 knobs to remove it from the console
8-7-16-2
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform "Functional Checks after FRU replacement (Debrief Guideline)" as described in
12 on page 8-223
NOTE:
Insert D Spring before assembling the knobs.
8 - 106
GE HealthCare CONFIDENTIAL
. 1
Manpower, Time, Tools, and PPE
Time
Approximately 5 minutes
Step
Section 8-7 - Replacing OPIO
LOGIQ T
Tools
No tools required
Corresponding graphic
Basic Service Manual
OTUS
8-2-6 "Tools needed
PPE
Cut resistant gloves