GE LOGIQ Totus Basic Service Manual page 613

Table of Contents

Advertisement

Direction 5936427, Rev. 1
10-4-6
Peripheral/option checks
If any peripherals or options are not part of the system configuration, the check can be omitted.
Table 10-5
Step
1
2
3
4
5
6
10-4-7
Mains cable inspection
Table 10-6
Step
1
2
3
4
5
10-4-8
Optional diagnostic checks
To complete the Ultrasound system checks, view the error logs and run desired diagnostics.
10-4-9
View the logs
1.) Review the Ultrasound system error log for any problems.
2.) Check the temperature log to if there are any trends that could cause problems in the future.
GE HealthCare CONFIDENTIAL
GE HealthCare Approved Peripheral/Hardware Option Functional Checks
Item
Media
Verify media drive(s) read/write properly. Clean if necessary.
B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
Color Printer
Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
ECG
Verify basic operation with customer
Footswitch
Verify that the footswitch is functioning as programed. Clean as necessary.
Mains Cable Inspection, As Appropriate
Item
Disconnect the mains cable from the wall.
Unplug cord
Disconnect the mains cable from the Ultrasound system.
Inspect
Inspect the mains cable and its connectors for damage of any kind.
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that
Verify
no strands may cause a short circuit.
Verify
Make sure the inlet connector retainer is functional.
Chapter 10 Care & Maintenance
LOGIQ Totus Basic Service Manual
Description
Description
10 - 11