8-7-14 Replacement of the T2 OPIO Lower Assy

This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
Table 8-50
Manpower
One person
8-7-14-1
Preparations
1.) Remove the OPIO Top Cover Set-T as described in
Set-T" on page 8-89
2.) Remove the UOPIO-T Assy as described in
8-98
8-7-14-2
Removal Procedure
Unscrew 4 screws (red circles) and disconnect the 1
1.
cable (blue circle) to remove the T2 OPIO Lower Assy.
8-7-14-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform "Functional Checks after FRU replacement (Debrief Guideline)" as described in
12 on page 8-223
8 - 104
GE HealthCare CONFIDENTIAL
. 1
Manpower, Time, Tools, and PPE
Time
Approximately 10 minutes
Step
Section 8-7 - Replacing OPIO
LOGIQ T
Tools
Philips Screwdriver
8-7-2 "Replacement of the OPIO Top Cover
8-7-10 "Replacement of the UOPIO-T Assy" on page
Corresponding graphic
Basic Service Manual
OTUS
8-2-6 "Tools needed
PPE
Cut resistant gloves
Knee Pad
Section 8-