Direction 5936427, Rev. 1
7-2-4
Collecting a screen capture with logs
If the LOGIQ Totus malfunctions, simultaneously press the Alt+D keys. Alt+D is available at all times
and collects a screen capture of the image monitor, user-defined presets, and these logs:
•
Keyboard Shadow Log (protected information)
•
Error Logs
•
Crash Log (protected information)
•
Vital Product Data
•
DICOM Logs
•
Windows Event Logs
•
Windows Modem Log
•
Diagnostic Logs
•
Service Logs
For a detailed list of Service Logs captured,
7-2-5 "Capturing network logs to use for troubleshooting" on page
7 - 4
GE HealthCare CONFIDENTIAL
Figure 7-1 ALT+D Dialog Box
Section 7-2 - Gathering Troubleshooting Data
LOGIQ Totus Basic Service Manual
7-8.