Direction 5936427, Rev. 1
7-3-5
Home (continued)
System Status
Use System Status to view status information on the LOGIQ Totus. Specifically, do the following:
•
View the hardware configuration file
•
View temperatures for FRUs and a graphical representation of the monitored temperatures
•
View statuses of the batteries
•
View voltages for FRUs.
•
View issues with the boot process
•
View alert logs
•
View logs that occurred before and after an error
•
View USB devices connected to the LOGIQ Totus
•
View the amount of time the LOGIQ Totus has been running
•
View the amount of time since the LOGIQ Totus has last been rebooted
The information on System Status is available to Class C and Class M licenses.
To access System Status, navigate to Utility (second page) > Service > Home.
GE HealthCare CONFIDENTIAL
Figure 7-10 System Status
Chapter 7 Diagnostics/Troubleshooting
LOGIQ Totus Basic Service Manual
7 - 19