GE LOGIQ Totus Basic Service Manual page 269

Table of Contents

Advertisement

Direction 5936427, Rev. 1
7-3-5
Home (continued)
System Status
Use System Status to view status information on the LOGIQ Totus. Specifically, do the following:
View the hardware configuration file
View temperatures for FRUs and a graphical representation of the monitored temperatures
View statuses of the batteries
View voltages for FRUs.
View issues with the boot process
View alert logs
View logs that occurred before and after an error
View USB devices connected to the LOGIQ Totus
View the amount of time the LOGIQ Totus has been running
View the amount of time since the LOGIQ Totus has last been rebooted
The information on System Status is available to Class C and Class M licenses.
To access System Status, navigate to Utility (second page) > Service > Home.
GE HealthCare CONFIDENTIAL
Figure 7-10 System Status
Chapter 7 Diagnostics/Troubleshooting
LOGIQ Totus Basic Service Manual
7 - 19