Replacement of the UAC-T PWA
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page
Table 8-89
Manpower
One person
8-9-2-1
Preparations
1.) Power off/Shutdown the system as described in
2.) Remove Unity Base Tray as described in
130
3.) Remove Front Cover Assy-T as described in
page 8-131
4.) Remove Unity Protector Assy as described in
page 8-133
8-9-2-2
Removal Procedure
1.
Unscrew 8 screws to remove the UAC&UCO-T Assy.
2.
Unscrew 9 screws to remove the UAC-T and Pull the
UAC-T PWA to remove the UAC-T PWA from the
UCO-T&UAC Assy.
Note: If the CWC-T is installed, remove the CWC-T as
described in
8-11-25 "Replacement of the CWC-T
PWA" on page 8-222
8-9-2-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform "Functional Checks after FRU replacement (Debrief Guideline)" as described in
12 on page 8-223
8 - 162
GE HealthCare CONFIDENTIAL
. 1
Manpower, Time, Tools, and PPE
Time
Approximately 15 minutes
Steps
Section 8-9 - Replacing Front End Parts
LOGIQ T
Tools
Phillips screwdriver
1-7-2 "Lockout/tagout procedure" on page 1-25
8-8-2 "Replacement of the Unity Base Tray" on page 8-
8-8-3 "Replacement of the Front Cover Assy-T" on
8-8-5 "Replacement of the Unity Protector Assy" on
Corresponding Graphic
Basic Service Manual
OTUS
8-2-6 "Tools needed
8-6.
PPE
Cut resistant gloves
Knee Pad
Section 8-