Connections on the patient I/O panel

Connect ECG
For information on the ECG, "Patient Cardiac and ECG Connections" in Chapter 3 of the Basic User
Manual (appropriate version).
3-5-6

Connecting probes

Probes can be connected at any time, whether the LOGIQ Totus is on or off.
DO NOT CONNECT A PROBE WHILE THE LOGIQ TOTUS IS POWERING UP.
CAUTION
The LOGIQ Totus has four DLP probe ports
LOGIQ Totus
Figure 3-7 Probe Connectors, Four DLP Probe Ports
To connect a probe:
NOTE:
It is not necessary to turn OFF power to connect or disconnect a probe. However, it is a good
idea to select a different probe or to freeze the image when removing a probe to avoid
disconnecting a live probe.
DO NOT ALLOW THE PROBE HEAD TO HANG FREELY. EXCESSIVE IMPACT TO THE PROBE
CAUTION
WILL RESULT IN IRREPARABLE DAMAGE.
KEEP THE PROBE CABLES AWAY FROM THE WHEELS.
CAUTION
DO NOT BEND THE PROBE CABLES.
DO NOT CROSS CABLES BETWEEN PROBES.
1.) Before connecting the probe:
a.) Perform a visual check of the probe and LOGIQ Totus sockets.
b.) Remove any dust or foam rests from the probe.
c.) Verify the probe and the probe cable for any visual damage.
2.) Hold the probe connector vertically with the cable pointing upward.
3.) Turn the connector locking handle counter-clockwise to the horizontal position.
4.) Align the connector with the probe port and carefully push into place.
5.) Turn the locking handle clockwise to the full vertical position to lock in place.
6.) Position the probe cable so that it is not resting on the floor.
3 - 16
GE HealthCare CONFIDENTIAL
(Figure
Section 3-5 - Completing the Setup
LOGIQ Totus Basic Service Manual
3-7). This is the only type of probe port on the