LOGIQ Totus Basic Service Manual
2-3-7
Environmental dangers (continued)
Patient Environment IEC60601-1 (IEC60601-1-1) and ANSI AAMI ES60601-1
Sub Clause 2.202 and figure 201 (IEC60601-1-1:2000)
Sub Clause 3.79 and figure A.9 (IEC60601-1:2005 and ANSI AAMI ES60601-1:2005)
Such an area is an environment in which medical diagnosis, monitoring or treatment is carried out. It is
very difficult to attach unique dimensions to the PATIENT ENVIROMENT.
In practice, a distance of 2,5 m (8.2 ft.) above the floor on which the medical personnel stand and a
horizontal distance of 1,5 m (4.9 ft.) have justified themselves as indicative of the dimensions of the
Patient Environment.
The patient environment/vicinity will be depicted as a dashed line in this procedure. example below.
Figure 2-4 Patient Environment
Item
Description
1.
Patient environment
2 - 16
Section 2-3 - Facility Needs