GE LOGIQ Totus Basic Service Manual page 533

Table of Contents

Advertisement

D
5936427, R
IRECTION
EV
8-11-17-2
Removal Procedure
1.
Push the button of Unity Protector Assy's connector
and disconnect the Unity Protector Assy's Connector
from the console.
2.
Unscrew 6 screws to remove the Unity UCO-T Front
Cover Brkt..
3.
Unscrew 2 screws to remove the U-DCWD Set from
the console.
8-11-17-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform "Functional Checks after FRU replacement (Debrief Guideline)" as described in
12 on page 8-223