D
5936427, R
IRECTION
EV
8-10-3-2
Removal Procedure
1.
Unscrew 10 screws to remove the Unity Rear
Shield Brkt from the console.
Note: If the WLAN-T was installed, remove the
WLAN-T as described in
of the WLAN-T and WLAN USB A Cable" on
page 8-199
Note: If the U-WIFI Module Assy-T was installed,
remove the U-WIFI Module Assy-T as described in
11-21 "Replacement of the U-WIFI Module
Assy" on page 8-214
2.
Unscrew 4 screws to remove the Unity Cable Cover
Brkt.
3.
Rotate the stud of HDMI cable(Main Monitor & Touch
panel) counterclockwise and disconnect from UBE-T
4.
Push the side of USB connector(OPIO1, OPIO2) and
disconnect from UBE-T.
NOTE: The connector can be inserted to any of
these two USB ports.
5.
Rotate the stud of D-sub connector counterclockwise
and disconnect from UBE-T.
8 - 170