General Console Requirements

2-2-1

Console Environmental Requirements

Table 2-2
Temperature
Humidity
Pressure
NOTE:
Temperature in degrees C. Conversion to Degrees F = (Degrees C * 9/5) + 32.
2-2-1-1
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and diameters can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interface.
2-2-2

Electrical Requirements

NOTE:
GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound
equipment. This dedicated power shall originate at the last distribution panel before the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet.
NOTE:
Please note that image artifacts can occur, if at any time within the facility, the ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
2 - 2
9
EVISION
Environmental Requirements for LOGIQ C Series Scanners
Operational
o
10 - 40
C
o
50 - 104
F
30 - 75%
non-condensing
700 - 1060hPa
Section 2-2 - General Console Requirements
LOGIQ C S
Storage
o
-5 - 50
C
-5 - 50
o
23 - 122
F
23 - 122
10 - 90%
10 - 90%
non-condensing
non-condensing
700 - 1060hPa
700 - 1060hPa
B
S
ERIES
ASIC
ERVICE
Transport
o
C
o
F