ECG display disappears after backup/restore

With ECG, after system auto-restart, turn system off and then back on. Otherwise ECG and TGC may
not function until next full power cycle.
7-7-2

Unable to measure in both stiffness and velocity in Breast and Thyroid

WIth Shearwave, when entering Measurement in Breast or Thyroid, only the stiffness measurements
show up on the touch panel if display units is set to kPa, even though both stiffness and velocity are
enabled in the Utility pages.