8-7-20 Replacement Of The Opio Mode Encoder Set-T - GE LOGIQ Totus Basic Service Manual

Table of Contents

Advertisement

D
5936427, R
IRECTION
EV

8-7-20 Replacement of the OPIO Mode Encoder Set-T

This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
Table 8-56
Manpower
One person
8-7-20-1
Preparations
1.) Remove the OPIO Top Cover Set-T as described in
Set-T" on page 8-89
2.) Remove the UOPIO-T Assy as described in
8-98
3.) Remove the T2 OPIO Upper Assy as described in
Assy" on page 8-105
4.) Remove the OPIO Laser Color Trackball with Cable as described in
Trackball assy-U_FLT20UFA-GJ" on page 8-101
5.) Remove OPIO Knob Set-T Assy as described in
page 8-106
6.) Remove OPIO Lower Cover Set as described in
Set-T" on page 8-107
7.) Remove OPIO Button Set-T as described in
the User Define Key Cap Set-T" on page 8-109
8.) Remove T2 OPIO Key Rubber as described in
on page 8-110
8 - 112
GE HealthCare CONFIDENTIAL
. 1
Manpower, Time, Tools, and PPE
Time
Approximately 15 minutes
Section 8-7 - Replacing OPIO
LOGIQ T
Tools
Philips Screwdriver
Spanner 11
8-7-2 "Replacement of the OPIO Top Cover
8-7-10 "Replacement of the UOPIO-T Assy" on page
8-7-15 "Replacement of the T2 OPIO Upper
8-7-16 "Replacement of the OPIO Knob Set-T" on
8-7-17 "Replacement of the OPIO Lower Cover
8-7-18 "Replacement of the OPIO Button Set-T and
8-7-19 "Replacement of the T2 OPIO Key Rubber"
Basic Service Manual
OTUS
8-2-6 "Tools needed
PPE
Cut resistant gloves
Knee Pad
8-7-11 "Replacement of the