D
5936427, R
IRECTION
EV
8-11-22 Replacement of the UMC-T
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page
Table 8-117 Manpower, Time, Tools, and PPE
Manpower
One person
8-11-22-1
Preparations
1.) Power off/Shutdown the system as described in
2.) Remove the Rear Cover Mid Assy-T as described in
Assy-T" on page 8-138
3.) Remove the Rear Cover Top Assy-U as described in
Assy-U" on page 8-143
8-11-22-2
Removal Procedure
1.
Unscrew 10 screws to remove the Unity Rear Shield
Brkt from the console.
Note: If the WLAN-T was installed, remove the
WLAN-T as described in
the WLAN-T and WLAN USB A Cable" on page
8-199
2.
Unscrew 4 screws to remove the UMC-T and Pull the
UMC for disconnecting from the UBE.
8-11-22-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform "Functional Checks after FRU replacement (Debrief Guideline)" as described in
12 on page 8-223
GE HealthCare CONFIDENTIAL
. 1
Time
Approximately 15 minutes
Steps
8-11-13 "Replacement of
Section 8-11 - Replacing Options
LOGIQ T
Tools
Phillips screwdriver
1-7-2 "Lockout/tagout procedure" on page 1-25
8-8-10 "Replacement of the Rear Cover Mid
8-8-15 "Replacement of the Rear Cover Top
Corresponding Graphic
Basic Service Manual
OTUS
8-2-6 "Tools needed
8-6.
PPE
Cut resistant gloves
Knee Pad
Section 8-
8 - 215