General Console Requirements

2-2-1

Console Environmental Requirements

2-2-1-1
If the unit is very cold or hot
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate to
CAUTION
its operating environment.
Table 2-2
DON'T STORE AT
THESE
TEMPERATURES
o
C
-40
-35
-30
-25
o
F
-40
-31
-22
-13
Hrs
20
18
16
14
2-2-1-2
Environmental specifications
Table 2-3
Operating temperature
o
o
10 to 35
C (50 to 95
F)
2-2-1-3
Cooling
The cooling requirement for the Vivid 7 is 3500 BTU/hr. This figure does not include cooling needed for
lights, people, or other equipment in the room. Each person in the room places an additional 300 BTU/
hr. demand on the cooling system.
2-2-1-4
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
2 - 2
11
Vivid 7 Acclimate Time
KEEP STORAGE TEMPERATURE WITHIN THESE LIMITS
-20
-15
-10
-5
0
-4
5
14
23
32
12
10
8
6
4
Environmental Specifications for Vivid 7 Scanners
Storage temperature
o
o
-20 to 60
C (-4 to 140
F)
Section 2-2 - General Console Requirements
5
10
15
20
25
41
50
59
68
77
2
0
0
0
0
Humidity
< 90% rH non-condensing
V
7 S
IVID
ERVICE
30
35
40
45
50
86
95
104
113
122
0
0
0
2
4
Heat dissipation
3500 BTU pr hour