Direction 5936427, Rev. 1
4-2-3

Power ON/Boot Up

Warnings
ALWAYS CONNECT THE LOGIQ TOTUS TO A FIXED POWER SOCKET WHICH
DANGER
HAS THE PROTECTIVE GROUNDING CONNECTOR.
NEVER USE A THREE-TO-TWO PRONG ADAPTER; THIS DEFEATS THE
DANGER
SAFETY GROUND.
ENSURE THAT THE POWER CORD AND PLUG ARE INTACT AND THAT THE
DANGER
POWER PLUG IS THE PROPER HOSPITAL-GRADE TYPE (WHERE REQUIRED).
LOGIQ TOTUS REQUIRES ALL COVERS.
CAUTION
ONLY OPERATE THE LOGIQ TOTUS WHEN ALL OF THE BOARD COVERS AND FRAME
PANELS ARE SECURELY IN PLACE. THE COVERS ARE REQUIRED FOR SAFE OPERATION,
GOOD PERFORMANCE, AND COOLING PURPOSES.
Use only power supply cords, cables, and plugs provided by or designated by GE HealthCare.
CAUTION
NOTE:
Do not cycle the circuit breaker ON-OFF-ON in less than five seconds. When turning OFF the
circuit breaker, WAIT until the ON/OFF button is no longer lit. The LOGIQ Totus should de-
energize completely before turning the circuit breaker ON.
GE HealthCare CONFIDENTIAL
Chapter 4 Functional Checks
LOGIQ Totus Basic Service Manual
4 - 7