Direction 5936427, Rev. 1
10-4-10
Physical inspection
NOTE:
These features may not be present on all Ultrasound systems.
Table 10-7
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
10 - 12
GE HealthCare CONFIDENTIAL
Physical Checks
Item
Labeling
Verify that all Ultrasound system labeling is present and in readable condition.
Scratches &
Inspect the exterior for dents, scratches or cracks.
Dents
Where applicable, verify that all covers are secured in place and are properly aligned with other
Covers
covers. Replace any covers that are damaged.
Input Power
Refer to:
10-4-7 "Mains cable inspection" on page
External I/O
Check all connectors for damage.
• Where applicable, check all wheels and casters for wear and verify the operation of foot brake,
Wheels and
which stops the Ultrasound system from moving, and the release mechanism.
Brakes
• Where applicable, check all wheel locks and wheel swivel locks for proper operation.
• Where applicable, verify the ease of Operator Panel (Operator Control Panel) movement in all
Control Panel
acceptable directions.
Movement
• Where applicable, ensure that the Control panel latches in position as required.
Control Panel
Check for proper operation of all operator panel lights.
Lighting
Inspect the displays for scratches and bad pixels.
Verify the proper operation of Contrast and Brightness controls.
Where applicable, confirm that the monitor arm allows
Display Monitor
• Swiveling the screen to the left and to the right.
• Folding the screen to the locked position.
• Can be adjusted in the up/down positions.
Cables and
Check all internal cable harnesses and connectors for wear and secure connector seating. Pay
Connectors
special attention to probe strain or bend reliefs.
Shielding and
Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Covers
Missing covers and hardware could cause EMI/RFI problems while scanning.
Control Panel &
Inspect alphanumeric keyboard and operator panel. Record any damaged or missing items.
Keyboard
(Replace faulty components as necessary).
Probe Holders
Where applicable, inspect the Probe Holders for cracks or damage.
Power and
System Status
Check for proper operation of all Power and System Status Indicators.
Indicators
Where applicable, check that the batteries are not damaged, do not leak, do not emit an odor, and
Batteries
are not deformed or discolored. Observe all warnings and cautions for battery handling, recharging,
storing, and/or disposal.
Section 10-4 - System Maintenance
LOGIQ Totus Basic Service Manual
Description
10-11.