Direction 5936427, Rev. 1
10-6-6-2
Generic Procedure
The testing meter is connected from accessible protectively-earthed parts of the equipment to ground.
Measurements should be made with power line polarity normal and reversed, and with the neutral open
and closed. Record the highest reading of current.
Figure 10-12 Set Up for Test of Earth Leakage Current, AAMI/ANSI ES 60601-1/IEC 60601-1 Clause 19
NOTE:
Color of Power outlet cable
i) USA and Canada
Hot:
Neutral: White
Ground: Green or Green-yellow
ii) Others
Hot:
Neutral: Blue
Ground: Green-yellow
When using Dale 600/601 or a similar test instrument, its power plug may be inserted into the wall outlet
and the equipment under test is plugged into the receptacle on the panel of the meter. This places the
meter in the grounding conductor and the current flowing through protective earth will be indicated in
any of the current ranges. The maximum allowable limit for earth leakage current is shown in Table 10-
10.
Table 10-15 Typical Data Sheet for Earth Leakage Current
Tester Neutral
10 - 30
GE HealthCare CONFIDENTIAL
Black
Brown
Tester Polarity Switch
Closed
Closed
Open
Open
Section 10-6 - Electrical Safety Tests
Enter Name of tested peripheral here:
Normal
Reversed
Normal
Reversed
LOGIQ Totus Basic Service Manual
Measured Earth leakage current