GE LOGIQ Totus Basic Service Manual page 554

Table of Contents

Advertisement

D
5936427, R
IRECTION
EV
8-12-45
Replacement of the OPIO Button Set-T and the User Define Key Cap Set-T
Functional Check
4-2-3 "Power ON/Boot Up"
4-3-7 "Probe/connector functional checks"
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
4-3-20 "Basic measurement functional checks"
4-2-4 "Power shut down"
8-12-46
Replacement of the T2 OPIO Key Rubber
Functional Check
4-2-3 "Power ON/Boot Up"
4-3-7 "Probe/connector functional checks"
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
4-3-20 "Basic measurement functional checks"
4-2-4 "Power shut down"
8-12-47
Replacement of the OPIO Mode Encoder Set-T
Functional Check
4-2-3 "Power ON/Boot Up"
4-3-7 "Probe/connector functional checks"
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
4-3-20 "Basic measurement functional checks"
4-2-4 "Power shut down"
8 - 228
Section 8-12 - Functional Checks after FRU replacement (Debrief Guideline)
GE HealthCare CONFIDENTIAL
. 1
Debrief
Script
Equipment passed all required checks and is ready for
use.
Debrief
Script
Equipment passed all required checks and is ready for
use.
Debrief
Script
Equipment passed all required checks and is ready for
use.
LOGIQ T
Basic Service Manual
OTUS