GE LOGIQ Totus Basic Service Manual page 405

Table of Contents

Advertisement

GE HealthCare CONFIDENTIAL
D
5936427, R
. 1
LOGIQ T
Basic Service Manual
IRECTION
EV
OTUS
8-6-3-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform "Functional Checks after FRU replacement (Debrief Guideline)" as described in
Section 8-
12 on page 8-223
NOTE:
Align and fix the white double lined mark on the UBE TO MONITOR HDMI CABLE and green
label attached on the OPIO HUB TO MONITOR POWER CABLE prior to second fixture using
CABLE TIE (2EA) as shown in the picture above.
- Fix the white double lined mark on the cable prior to second fixture using cable tie(2EA)
- Fix the green label attached on the cable prior to second fixture using cable tie(2EA)
Section 8-6 - Replacing Monitor and Monitor Arm
8 - 79