Direction 5936427, Rev. 1
3-7-9-4
Dataflow Settings
Use the Dataflow Settings page to set parameters for dataflows.
Table 3-39 To Configure Dataflow Settings
1.
Navigate to Connectivity > Connection Manager > Advanced Settings.
2.
Select Dataflow Settings.
3.
Under Automatic Disable Patient Data, select to automatically disable patient data.
NOTE: If selected, locks the patient name, date of birth and gender (like Patient ID). The Factory Default for this preset is disabled.
4.
To Archives patient data automatically, enable Auto Archiving Patient Data.
5.
Under After [End Current Patient], go to:, select Worklist screen or Patient screen.
6.
Under Double click on patient list to start, select Review or New Exam to display each time you double click on the patient name in the
patient list on the Patient menu.
GE HealthCare CONFIDENTIAL
Step
Chapter 3 LOGIQ Totus Setup
LOGIQ Totus Basic Service Manual
Corresponding Graphic
3 - 77