Required Adjustment after Replacing Assembly
After replacing the following assembly, the adjustment items described in Table 3-1 must
be required. The adjustment must be performed Agilent's qualified service personnel. If
you need the adjustment for your E5052A, it should be sent to the nearest Agilent
Technologies service office.
Table 3-1
Required Adjustment Item after Replacing Assembly
Replaced Assembly
Adjustment Item
A1 Freq/Power Det.
√
A2 DC Source
A3 ADC/DSP
A4/A5 Local Synt
A6 Phase Det
√
HDD
√
A21 PCI DSP Card
Writing ID
This item writes the serial number and the option structure into the E5052A.
Required equipment for the writing ID
None
DC Control Voltage Adjustment
The purpose of this procedure is to adjust the DC control voltage of the DC source of the
A2 DC Source board.
Chapter 3
Required Adjustment after Replacing Assembly
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
Adjustment
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
75