Required Adjustment after Replacing Assembly

After replacing the following assembly, the adjustment items described in Table 3-1 must
be required. The adjustment must be performed Agilent's qualified service personnel. If
you need the adjustment for your E5071C, it should be sent to the nearest Agilent
Technologies service office.
Table 3-1
Required Adjustment Item after Replacing Assembly
Replaced Assembly
OCXO Board
Analog Base Module
Synthesizer Module for Source
Synthesizer Module for Local
Level Vernier Module
Receiver Module
A21
Distributor Module
RF Switch

Writing ID

This item writes the serial number and the option structure into the E5071C.
Required equipment for the writing ID
None

OCXO Adjustment

The purpose of this procedure is to adjust the 10 MHz OCXO.
Chapter 3
Required Adjustment after Replacing Assembly
Adjustment Item
Adjustment