Storing, Logging and Printing
Saving Graphics Results to Instrument Store
Saving Graphics Results to Instrument Store
Description
Graphical representation of measurement results is very useful particularly during a
long measurement period. It provides an overview of the results and can be printed
for record keeping.
Graphics results can be stored in instrument graph storage or on floppy disk.
1 Before starting your measurement, choose the GRAPH STORAGE resolution
HOW TO:
and location.
The resolution chosen affects the ZOOM capability when viewing the bar graphs.
If 1 MIN is selected, 1 MIN/BAR, 15 MINS/BAR and 60 MINS/BAR are
available. If 15 MINS is selected, 15 MINS/BAR and 60 MINS/BAR are
available. If 1 HOUR is selected, 60 MINS/BAR is available.
The graphics results can be stored in the instrument - INTERNAL or stored on
DISK. Storage to disk will use a default file name unless a file name is specified
on the
page 290.
2 Press
the chosen location.
244
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
OTHER
display. See "Saving Graphical Results to Disk"
FLOPPY DISK
to start the measurement. Graphical results will be stored in
RUN/STOP