ATM Operation
ATM Service Disruption
Background Traffic " page 161 & "Setting Foreground and Background
Distributions " page 162).
3
Set up the Receiver for ATM. (See "Setting up the Receiver for ATM Payloads
" page 170).
4 In the Receiver ATM SETTINGS folder set up the service and filter required.
5 Verify that no errors/alarms are received (no red LEDs and the RESULTS page
is clear of errors).
6 In the system under test, activate the protection switch or generate some kind of
temporary system failure.
7 Go to the RESULTS page. In the first RESULTS field select
and note the duration of the service disruption. Available measurements are
LONGEST burst, SHORTEST burst and LAST burst.