Selecting Test Features
Adding Pointer Adjustments
G.783(g) 87-3 Pattern
Pointer Adjustment
Pointer Sequence
G.783(a) PERIODIC
Periodic Single adjustments, each with opposite polarity to the preceding
SINGLE
adjustment. The interval between pointer adjustments is user selectable (see
Note 1 page 93).
G.783 (b) PERIODIC
Periodic Single adjustments, with selectable polarity and added adjustment (1
ADD
extra). The spacing between the added adjustment and the previous adjustment
is set to the minimum, (see Note 2 page 93). The interval between pointer
adjustments is user selectable (see Note 1). Added adjustments occur every 30
seconds.
G.783 (c) PERIODIC
Periodic Single adjustments, with selectable polarity and cancelled adjustment
CANCEL
(1 less). The interval between pointer adjustments is user selectable (see Note
1 page 93). Cancelled adjustments occur every 30 seconds.
G.783(d) PERIODIC
Periodic Double adjustments (pair of adjustments). The pair alternate in
DOUBLE
polarity. The spacing between pairs of adjustments, of like polarity is set to the
minimum (see Note 2). The interval between pointer adjustments is user
selectable (see Note 1).
G.783 (e) SINGLE
Periodic Single adjustments, all of the same polarity which is selectable.
Separation between pointer adjustments is fixed at approximately 30 seconds.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
No Pointer
Adjustment
87
An Example of a Pointer Sequence
Description
Start of Next
87-3 Pattern
3