Making Measurements
Measuring Jitter Transfer
12 With MODE set to
The Jitter Transfer display is replaced by an information display for the duration
of the Calibration.
A bar graph showing the progress of the calibration will appear on the display.
When the Calibration is complete, the display will revert to the
HOW TO:
Start the Jitter Transfer Measurement
N O T E
The Jitter Transfer measurement must be started within 10 minutes of the completion
of calibration.
1 After the CALIBRATION is completed, remove the back to back connection
from the PDH or SDH or optical interfaces.
If the measurement is to be made at a network equipment monitor point, choose
the jitter transfer measurement.
2 Choose MODE
The measurement's progress can be monitored on the
HOW TO:
View the Results
1
Select
FORMAT.
146