Agilent Technologies OmniBER 718 SDH User Manual page 224

Table of Contents

Advertisement

Packet over SDH (POS)
Jitter Tolerance Testing of POS Equipment
8 Select the
display above).
9
Choose the NUMBER OF POINTS at which jitter is transmitted (3 to 55)
10 Choose the DWELL TIME - the time jitter is generated at each jitter frequency
point (0.1 to 99.9 seconds).
11 Choose the DELAY TIME - the time delay between the jitter frequency/
amplitude being applied and the error measurement being made. This allows the
network equipment to settle as jitter frequency is changed. (0.1 to 99.9 seconds).
12 Set the ERROR THRESHOLD field to ANY ERRORS. Any BIT, BIP, FCS or
header errors will result in a FAIL.
13 Press
The measurements progress can be monitored on the
end of the test the results can be viewed on the
displays. The
the results remain on the
measurement is made.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
folder and select AUTO TOLERANCE and a MASK (see
JITTER
to start the jitter auto tolerance measurement.
RUN/STOP
display is cleared when
TRANSMIT
RESULTS
display. At the
TRANSMIT
or
RESULTS
TRANSMIT
is pressed but
TRANSMIT
display until the next jitter tolerance