Back End Processor (Bep) Parts Replacement; Back End Processor (Bep) Replacement - GE VIVID E9 Service Manual

Hide thumbs Also See for VIVID E9:
Table of Contents

Advertisement

GE H
EALTHCARE
D
GA091568, R
IRECTION
Section 8-11

Back End Processor (BEP) parts replacement

8-11-1

Back End Processor (BEP) replacement

If the old BEP is a BEP5 with 4D, and the new BEP is a BEP6, a new Graphics Adapter is needed.
If present, the DVR Board must be moved over from the old to the new BEP.
8-11-1-1
Warnings
ENERGY CONTROL AND POWER LOCKOUT FOR VIVID E9.
WARNING
WARNING
WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO
VOLTAGE GREATER THAN 30 VOLTS:
1. FOLLOW LOCK OUT/TAG OUT PROCEDURES.
TAG
&
LOCKOUT
2. TURN OFF THE BREAKER.
Signed
Date
3. UNPLUG THE VIVID E9.
4. MAINTAIN CONTROL OF THE POWER PLUG.
5. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE, AS THERE ARE NO TEST POINTS TO
VERIFY ISOLATION. THE AMBER LIGHT ON THE OP PANEL ON/OFF BUTTON WILL TURN OFF.
BEWARE THAT THE MAIN POWER SUPPLY AND BACK END PROCESSOR MAY BE ENERGIZED EVEN IF THE POWER
IS TURNED OFF WHEN THE CORD IS STILL PLUGGED INTO THE AC OUTLET.
WHEN YOU RETURN THE USED BEP TO YOUR LOCAL PARTS ORGANIZATION, MAKE
WARNING
WARNING
SURE YOU REMOVE ALL PATIENT DATA FROM THE HARD DRIVE, GIVEN THAT THE
HARD DRIVE IS STILL FUNCTIONAL.
IN SOME COUNTRIES, YOU MAY BE REQUIRED TO DELETE ALL SOFTWARE FROM
THE DISK BEFORE RETURNING THE BEP TO THE PARTS WAREHOUSE. FOLLOW
YOUR LOCAL POLICIES.
BEFORE YOU DISPOSE OF THE HARD DRIVE, MAKE SURE YOU REMOVE ALL
WARNING
WARNING
PATIENT DATA FROM THE HARD DRIVE, GIVEN THAT THE HARD DRIVE IS STILL
FUNCTIONAL.
IN SOME COUNTRIES, YOU MAY BE REQUIRED TO DELETE ALL SOFTWARE FROM
THE DISK BEFORE RETURNING THE HARD DRIVE TO THE PARTS WAREHOUSE.
FOLLOW YOUR LOCAL POLICIES.
8 - 244
5
EVISION
Section 8-11 - Back End Processor (BEP) parts replacement
VIVID E9 S
M
ERVICE
ANUAL