Updating The Device Firmware - Thermo Scientific VF-F20-A Operating Manual

Integral fraction collector ft
Table of Contents

Advertisement

7.15 Updating the Device Firmware

When
Items required
Preparations
Follow these steps
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Updating the device firmware might be required, for example, when a
new firmware version is released that adds functionality or solves
problems of a previous version.
Firmware version/Chromeleon Service Release as appropriate
TIP When a new firmware version is released, the new version will be
included in the next available Chromeleon Service Release. The new
firmware will not be transferred automatically to the device when you
install the Chromeleon Service Release.
1. cacl.
2. Connect the device in the Chromeleon software.
3. Stop all operations in the Instrument that includes the device.
4. Wait until the instrument is idle.
1. Start the Chromeleon Instrument Configuration Manager.
2. Perform a firmware update from the General tab page in the
configuration dialog box for the device. For more information, refer
to the Chromeleon Help.
7
Maintenance and Service
Page 189