Important Operating Parameters

Integral Fraction Collector FT (VF-F20-A) Operating Manual
The parameters described in this section should be considered for
routine operation of the device. You can usually access these
parameters from the Chromeleon user interface.
If a parameter listed below is not available in Chromeleon, consider
updating the firmware and Chromeleon version.
For more information, refer to Chromeleon Help and documents.
Parameters
Rack type (blue, green, red, yellow)
Change rack
Tube position
Reset start
position
Temperature control
Temperature
nominal
Description
The rack type specifies the type of sample
rack or well plate that is installed in the
respective segment (RackType_Blue, _Green,
_Red or _Yellow).
For details on the rack type settings, see
Type Settings
(
page
111).
Rotate the required segment of the carousel
to the front to install or remove a sample rack
or well plate.
The tube position identifies the position into
which the next fraction is collected.
The Tube Position consists of the color code
for the segment and the position on the
sample rack or well plate, which are
separated by a colon (for example, B:E8 for
segment B, sample position E8).
Set the start position to the first available
position of the carousel. The segments are
checked in Red, Yellow, Blue, and Green
order to determine the start position. For
example, R:A1 should be the new position if
there is a rack or well plate on Red segment.
Enable or disable the temperature control for
the sample compartment. As a standard, the
temperature control is disabled.
The temperature setting defines the setpoint
to which the fraction collector compartment
is to be cooled or heated. The default
temperature is set to 25 °C (Temperature
Nominal = 25).
For details, see
Thermostatting the Fraction
Collector Compartment
(