Connecting the Flush Solvent Reservoir
Integral Fraction Collector FT (VF-F20-A) Operating Manual
1. Connect the flush solvent reservoir to the line:
a) Rinse the flush solvent reservoir thoroughly with a high-purity
solvent.
b) Fill the flush solvent reservoir with fresh flush solvent.
c) Feed the flush solvent tubing through the 1/16'' retaining guide
and through an open hole in the cap of the flush solvent
reservoir. The retaining guide keeps the tubing in place in the
reservoir. Close any open holes in the reservoir cap with cap
plugs.
Figure 35: Securing the flush solvent tubing with the reservoir cap
No.
Description
1
Flush solvent tubing (OD 1/16'')
2
Retaining guide, 1/16''
3
Reservoir cap
4
Cap plugs
2. Close any open holes in the reservoir cap with cap plugs.
3. Slide the filter holder and adapter onto the end of the flush solvent
tubing.
4. Tighten the reservoir cap hand-tight. Press the retaining guide into
the hole in the reservoir cap to ensure that the tubing is kept in
place in the cap.
5. Place the flush solvent reservoir in the solvent rack. Position the
solvent lines straight in the flush solvent reservoir and tubing guides.
5
Installation
Page 89