Troubleshooting
Firmware Failure
Page 204
Chromeleon: The Instrument Audit Trails can be found on the ePanel
Audit tab. In addition, Audit Trails for each instrument are available in
the Chromeleon Console Data view in the folder of the instrument.
Messages in the Chromeleon Audit Trail are preceded by an icon. The
icon identifies the seriousness of the problem. The table shows the icons
and explains the severity level.
Icon
Severity
Warning
Error
Abort
or
For possible causes and recommended remedial actions, see
(
page
205).
If a firmware failure occurred during operation of the device, an
exception log was created about the processes during the firmware
failure. The firmware sends the exception log to the Audit Trail when the
device is connected in Chromeleon software.
In this case, observe the following:
Send the daily Audit Trail as .cmbx file to Technical Support before
you clear the log.
To clear the exception log and continue operation of the device,
perform the command ExceptionLogClear.
For more information, refer to the Chromeleon Help.
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Description
The queue can be started.
The current run is not interrupted.
However, Thermo Fisher Scientific recommends taking
appropriate action to remedy the situation.
The software attempts to correct the problem. An error
does not interrupt the current analysis. However, if the
error occurs during the Ready Check, the analysis will not
be started.
The queue (cannot be started.
A running queue or batch is stopped immediately.
The issue must be resolved first.
Messages