Operating Issues; Resolving Front Leakage - Thermo Scientific VF-F20-A Operating Manual

Integral fraction collector ft
Table of Contents

Advertisement

8
Troubleshooting
8.3

Operating Issues

8.3.1

Resolving Front Leakage

When
Parts required
Additional items required
Preparations
Follow these steps
Page 212
This section gives an overview of possible operating issues and remedial
actions.
The leak sensor is wet. The leak sensor reports leakage.
Replacement part as required
Cloth or tissue
When resolving leakage, observe the safety guidelines and general rules
for maintenance and service as presented in
(
page
142).
1. Locate the source of the leak.
As leakage usually occurs at a connection, visually inspect all
components and connections in the flow path.
2. Tighten or replace the connection or component as required.
3. With a cloth or tissue, thoroughly absorb all liquid that has collected
in the leak tray and under the leak sensor. Be careful not to bend the
sensor.
4. Allow the sensor to adjust to the ambient temperature for a few
minutes.
5. If leakage is no longer reported, you can resume operation.
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Maintenance and Service