Maintenance and Service
Page 180
4. Check if the needle unit is positioned correctly:
Move the lock pin slightly upward. If the lock pin and the vial pusher
can move up and down easily, the needle unit sits correctly in place.
5. Reinstall the needle capillary. See
(
page
172).
6. Install the front plate. See
7. Power on the device. Close the compartment doors and the device
will run initialization automatically. Make sure no errors or warnings
are reported in the Audit trail.
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Replacing the Needle Capillary
Removing the Front Plate
(
page
155).