Thermo Scientific VF-F20-A Operating Manual page 118

Integral fraction collector ft
Table of Contents

Advertisement

6
Operation
Page 118
Parameters
Delay volume
Delay capillary
IDxL
Delay capillary
ID
Delay capillary L
Fraction collection options
Off (do not
collect at all)
Collect by peak
Collect by time
Tube wrapping
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Description
The Fraction Delay Volume is the fluidic
volume travelled by a peak after triggering
the set start threshold conditions between a
detector and the separation point of the
downstream fraction collection valve.
The volume can be logged automatically if the
Vanquish Fraction Collector Delay Volume
Determination operational qualification is
performed. Before running the OQ test, it is
recommended to perform first an Instrument
Installation Qualification to ensure that the
standard HPLC system is basically ready to
operate in its intended use.
For details, see
Delay Volume Determination
(DVD)
(
page 120).
The Delay Volume can be changed by
specifying delay Capillary inner diameter (ID)
and length (L).
Select either Custom or predefined Delay
Capillary IDxL. If the predefined Delay
Capillary IDxL is selected, the Delay Volume
is automatically updated.
If Custom as Delay Tubing Capillary IDxL is
selected, then need to manually input Delay
Capillary ID and Delay Capillary L. The Delay
Volume is calculated and displayed
accordingly.
Select this option to disable fraction
collection.
When this option is selected, fraction
collection starts depending on the specified
detection channels and peak detection
settings.
Fraction collection is independent of peak
detection.
When this option is selected, use the
Collection period input field to specify the
time in [s] after which a new tube will be
used.
Select this check box to return to the first
tube after the last tube is filled. If this check
box is cleared, the Chromeleon Queue will be
aborted when the last tube is reached.