7
Maintenance and Service
Follow these steps
Page 200
1. Unpack the slide-in module. Remove any foam spacers from the
module.
2. Push all tubing and capillaries, which are present in the tubing chase
of the Vanquish system modules, into the tubing chase. Otherwise,
you will not be able to insert the slide-in module properly into the
enclosure in the next step.
3. Insert the slide-in module in the enclosure.
The following steps require a team effort:
a) Take the slide-in module on both sides from below.
b) Lift the slide-in module to the height of the enclosure.
c) Place the slide-in module in the enclosure so that the module
sits in the enclosure by approximately 25 cm.
d) Push the slide-in module onto the rails and into the enclosure
until the slide-in module sits completely in the enclosure.
Figure 87: Inserting the slide-in module
4. Gradually and evenly, tighten the four captive screws on the slide-in
module hand-tight.
NOTICE Verify that the screws are tightened. Pull the slide-in module
by the leak tray towards the front and check whether the screws
move. If they do not move, the slide-in module is installed properly.
If the screws move, tighten the screws further.
Integral Fraction Collector FT (VF-F20-A) Operating Manual