5
Installation
5.9

Flush System

5.9.1

Choosing the Flush Solvent

5.9.2

Connecting Flush Solvent Reservoir

Parts required
Preparations
Page 86
Use the flush solvent that fulfills the following requirements:
Before filling the flush solvent reservoir, rinse the reservoir
thoroughly. Make sure that no particles, dust or algae are present.
Recommended when using 100% water as flush solvent
Replace the flush solvent daily.
Use a flush solvent that is suitable for your application and that
removes residual fraction from the flush buffer loop sufficiently.
Flush solvent reservoir
Flush solvent tubing
Retaining guide, 1/16''
Solvent line filter
Filter frit
Solvent line adaptor (1/8'' - 1/16'')
Reservoir cap
Cap plugs
1. Prepare the flush solvent and flush solvent reservoir. Observe the
flush solvent guidelines in
2. Assemble the filter holder and solvent line adapter, wearing
appropriate clean gloves.
a) Place the frit in the filter holder (bottom part).
b) Make sure that the frit is in a level position.
c) Screw the filter top to the filter bottom.
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Choosing the Flush Solvent
(
page
86).