Thermo Scientific VF-F20-A Operating Manual page 112

Integral fraction collector ft
Table of Contents

Advertisement

6
Operation
Page 112
Below table is listed with auto needle height default values and settable
ranges for different rack types that are provided by Thermo Scientific.
Rack Type
ThermoVial54
ThermoVial96_6mm
ThermoVial96_7mm
ThermoVial96_8mm
ThermoVial9
ThermoVial16
WellPlate96
WellPlate384
For AboveVial mode the default needle height value considers a
distance of 2 mm between the height of the vial and the needle tip to
ensure no needle obstruction can happen.
For the use of well plates in AboveVial mode the plate height must be
configured into the system manually. For details of needle positioning
mode settings, refer to the Chromeleon Help.
NOTICE
It is highly recommended to use the Vanquish rack type barcoded
sample racks and well plates. Following actions may cause needle
damage, sample loss and solvent spraying:
Using unknown racks or well plates
Not properly setting the needle height parameters
Not properly placing the racks or well plates in fraction collector
compartment.
Integral Fraction Collector FT (VF-F20-A) Operating Manual
AboveVial Mode
Default
Needle
Needle
Height
Height
settable
[mm]
range [mm]
34
10-50
34
10-50
42
10-50
42
10-50
48
10-50
48
10-50
n.a.
10-50
n.a.
10-50
InVial Mode
Default
Needle
Needle
Height
Height
settable
[mm]
range [mm]
10
10-45
10
10-45
10
10-45
10
10-45
10
10-45
10
10-45
3
3-41.5
3
3-41.5