Thermo Scientific VF-F20-A Operating Manual page 208

Integral fraction collector ft
Table of Contents

Advertisement

8
Troubleshooting
Message and Code
Code 9012. Needle hit an unexpected
object. (Z-axis)
Code 9016. Flush pump stepper motor
over current.
Code 9020. TEA driver abnormal:
Heating current abnormal , Cooling
current abnormal, or IC fault feedback.
Code 9021. Temperature sensor open -
cabin air
Code 9022. Temperature sensor short -
cabin air
Code 9023. Temperature sensor open -
ambient air
Code 9024. Temperature sensor short -
ambient air
Code 9025. Temperature sensor open -
outside heat sink
Code 9026. Temperature sensor short -
outside heat sink
Code 9027. Cabin air temperature too
low
Code 9028. Cabin air temperature too
high
Code 9030. Temperature control is
switched off due to left door is open for
20 minutes.
Code 9031. Internal Fan1 broken
Code 9032. Internal Fan2 broken
Page 208
Description and Remedial Action
Needle hit an unexpected object due to incorrect sample rack or well
plate configuration, or improper needle height or well plate height is
set manually. Needle obstruction, maybe due to incorrect sample rack
or well plate configuration or any unexpected object. Check if the
correct sample rack or well plate is configured. Consider number of
positions of the configured sample rack or well plate as well as the
respective well plate height. Or check any unexpected object. After
resolving physical blockage, then run self-test to recover the issue.
Check if the correct sample rack or well plate are configured, and the
correct tube position is defined in the ePanel FC tab.
Check if the needle height or well plate height is set correctly.
If the issue persists, contact Technical Support.
Power cycle the device to recheck the status.
If the issue persists, contact Technical Support.
First run self-test to recover the issue and check audit trail for more
information. If the issue persists, power cycle the device and check
again if it is recovered.
If the issue still persists, contact Technical Support.
Power cycle the device .
If the issue persists, contact Technical Support.
Power cycle the device .
If the issue persists, contact Technical Support.
Power cycle the device.
If the issue persists, contact Technical Support.
Power cycle the device.
If the issue persists, contact Technical Support.
Power cycle the device.
If the issue persists, contact Technical Support.
Power cycle the device.
If the issue persists, contact Technical Support.
Cabin air temperature < 2°C
Power cycle the device. If the issue persists, contact Technical Support.
Cabin air temperature > 45°C
Power cycle the device. If the issue persists, contact Technical Support.
Check the state of the left door.
If the left door is forgotten to be closed, close the door.
Power cycle the device. If the issue persists, contact Technical Support.
Power cycle the device. If the issue persists, contact Technical Support.
Integral Fraction Collector FT (VF-F20-A) Operating Manual