When to Perform DVD Function

Integral Fraction Collector FT (VF-F20-A) Operating Manual
No.
Description
1
Delay capillary
2
Fraction collection valve
When collection is triggered during a run, the fraction collection valve is
switched from waste position to collection position using the following
delay time calculations:
Peak-based:
Start of fraction collection: t = t
End of fraction collection: t = t
Time-based:
Start of fraction collection: t = t
End of fraction collection: t = t
It is necessary to re-measure the delay volume if any of the below
conditions are met.
Initial installation of a new HPLC system with an Integral Fraction
Collector (VF-F20-A)
Replacing a detector flow cell
Replacing a delay capillary
Replacing a fraction collection valve
Replacing an air bubble sensor
Important!
Only qualified Thermo Fisher engineer can replace the fraction
collection valve and the air bubble sensor.
Never move or uninstall the fitting on port 3 of fraction collection
valve. The flow path on this port have been performed calibration
during in-house production process or in-field service.
No.
Description
3
Air bubble sensor
+ Delay time
peak start
+ Delay time
peak end
+ Delay time
collection time window start
+ Delay time
collection time window end
6
Operation
Page 121