Performing A Flush Purge Cycle - Thermo Scientific VF-F20-A Operating Manual

Integral fraction collector ft
Table of Contents

Advertisement

7.8.3

Performing A Flush Purge Cycle

When
Preparations
Follow the step
Integral Fraction Collector FT (VF-F20-A) Operating Manual
13. In Chromeleon, make sure to check the installed flush buffer loop
compatible with the installed needle capillary. Both the needle
capillary and flush buffer loop shall match with the application of
flow rate range.
Flow rate [mL/min]
0 - 5.0
5.0 - 10.0
NOTICE The needle capillary must match with installed flush buffer loop
as shown in above table in order to achieve good fraction collection
performance with low carry over and high recovery when flush function
is enabled. The fluidic volume of flush buffer loop shall be around
double volume of matched needle capillary.
14. Flush the flush pump solvent lines to ensure they are filled with
fresh flush solvent, see
Routine and preventive maintenance for the flush purge system
Replacing the flush solvent
Replacing flush purge lines
Observe the Flush Solvent Guidelines. See
(
page
164).
In Chromeleon, select Purge Flush Pump on the ePanel FC tab to fill the
flush pump with the fresh flush solvent. During purging, the flush pump
and the flush buffer loop is washed continuously until all flush purge
lines filled with fresh flush solvent.
7
Maintenance and Service
Needle Capillary ID x L
[mm]
0.18 × 415
0.25 × 415
Performing A Flush Purge Cycle
Flush Solvent Guidelines
Flush Buffer Loop ID x L
[mm]
0.5 × 255 (V = 50 µL)
0.5 × 510 (V = 100 µL)
(
Page 171).
Page 171