Leak Detection - Thermo Scientific VF-F20-A Operating Manual

Integral fraction collector ft
Table of Contents

Advertisement

3
Device Overview
3.5

Leak Detection

Page 46
Leaks are a potential safety issue.
The front leak sensor monitors for liquid leaks from the flow
connections. The liquid is collected in the front leak tray and guided to
the drain port. From the drain port, the liquid is guided to waste through
the drain system of the Vanquish system.
When the front leak sensor detects leakage, the status indicators change
to red and beeping starts to alert you. Follow the instructions in this
manual to find and eliminate the source for the leakage. For more
details, see
Resolving Front Leakage
Integral Fraction Collector FT (VF-F20-A) Operating Manual
(
page
212).