5
Installation
Guide Holes and Capillary Clips
Page 70
Figure 22: Tubing chase with tubing guides (left: view from inside, right:
view from top)
No.
Use for
1
Solvent tubing (up to three solvent lines, preferably routed to the upper
degas chambers)
2
Solvent tubing (up to three solvent lines)
3
Wash liquid tubing (Autosampler seal wash, fraction collector needle wash)
4
Detector and fraction collector waste lines
Tubing brackets are available for holding the tubing in place. Slip the
bracket side onto the drain pipe.
Figure 23: Tubing bracket (left), tubing bracket installed (right)
Guide holes and capillary clips are provided at specific positions on the
system modules. Route flow connections from one module to the next
module in the Vanquish system through the appropriate guide hole or
capillary clip when instructed to do so in the manual.
Integral Fraction Collector FT (VF-F20-A) Operating Manual