How to Use the Tool
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Figure 47: Vanquish Fraction Collector Delay Volume Calculator
1. Select the Detector configured in the Vanquish system from the
drop-down list.
2. Select the Flow Cell Type installed in the detector from the drop-
down list.
3. Select the installed Delay Capillary IDxL from the drop-down list.
This will automatically fill the Delay Capillary Diameter and Delay
Capillary Length.
If Custom is selected as Delay Capillary IDxL the values for Delay
Capillary Diameter and Delay Capillary Length need to be entered
manually.
4. Define the Pump Flow Rate to be applied in the instrument
method.
5. Define the Detector Response Time to be applied in the
instrument method.
6. The calculator displays the resulting Delay Time and Delay
Volume in the Method Specific Delay section.
6
Operation
Page 129